Setting the readout, 3 setting the readout, Setting averaging level – Ohaus Voyager Balances User Manual

Page 127: Setting stability level, Setting auto-zero, Setting legal for trade

Advertising
background image

63

7.3 Setting the readout

This menu option enables you to set the balance averaging level, stability level, auto zero, and legal for trade (LFT).

Setting averaging level

Press Enter button, AVERAGING LEVEL FILTER is displayed. Averaging level
filter compensates for vibration or excessive air currents. A zero setting
(GOOD) corresponds to minimum filtering with fastest stabilization time. A
setting of (BEST)10 is maximum filtering with the slowest stabilization time.
Adjust to desired setting and press Enter button.

Setting stability level

Scroll to STABILITY LEVEL and press Enter button, STABILITY LEVEL FILTER is
displayed. Stability level filter determines the variation range in divisions for
a given reading depending upon the filter setting. A setting of zero (GOOD)
permits the balance to respond to variations of .5 divisions. A setting of
(BEST) 10 sets the threshold to 3 divisions. Adjust to desired setting and
press Enter button.

READOUT SETUP

AVERAGING LEVEL FILTER

GOOD

BEST

5

or

READOUT SETUP

STABILITY LEVEL FILTER

GOOD

BEST

5

or

Setting auto-zero

Scroll to AUTO ZERO and press Enter button, AZT is displayed. Auto zero
minimizes the effects of temperature changes and shift on the zero reading.
There are 4 settings, OFF, 0.5, 1, and 3. The numbers represent display
divisions. The balance maintains the zero display until the set threshold is
exceeded. Select the desired setting and press Enter button.

READOUT SETUP

AZT

or

OFF

0.5

1

3

Setting legal for trade

Scroll to LEGAL FOR TRADE and press Enter button. Legal for trade can be
set ON or OFF. When set ON, only the functions allowed by the national
weights and measures legislation are available. Press Enter button after
making your selection.

READOUT SETUP

AVERAGING LEVEL 5

STABILITY LEVEL 5

AUTO ZERO 0.5

LEGAL FOR TRADE OFF

EXIT TO WEIGH

or

SET BALANCE

Enter

From main menu

Select SET BALANCE and press the Enter button. Display advances to SET
BALANCE with READOUT highlighted. Press the Enter button, display
advances to READOUT SETUP.

READOUT

Balance settings