Span calibration printout, Linearity calibration printout, Calibration test printout – Ohaus Voyager Balances User Manual

Page 230

Advertising
background image

77

RS232 Interface/printing data

Span calibration printout

When performing a Span calibration with GLP turned ON, a printout is
automatically made after the calibration mass is placed on the pan.

- - - - - SPAN CAL - - - - - -

12/01/99

1:00:00 PM

Bal Id 1234
Cal:

400.000g

Old:

400.000g

Dif:

0.000g

Wt. Ref...................................... USER
NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

Linearity calibration printout

When performing a Linearity calibration with GLP turned on, a printout is
automatically made after the calibration mass is placed on the pan.

- - - - - LIN CAL - - - - - -

12/01/99

1:00:00 PM

Bal Id 1234
Cal:

400.000g

Old:

399.094g

Dif:

0.006g

Wt. Ref...................................... USER
NO 2056853
PROJ NO 100012
Name........................................

- - - - - END - - - - -

Calibration test printout

When performing a Calibration Test with GLP turned on, a printout is
automatically made after the calibration mass is placed on the pan.

- - - - - CAL TEST - - - - - -

12/01/99

1:00:00 PM

Bal Id 1234
Cal:

400.000g

Act:

400.004g

Dif:

0.004g

Wt. Ref......................................
USER NO 2056853
PROJ NO 100012
Name........................................
- - - - - END - - - - -