Enter oil density, Testing the sample, Auto sample – Ohaus Voyager Balances User Manual

Page 151: Auto print, Enter

Advertising
background image

87

WEIGH IN AIR: 207.457

TEMP: 25.0

WEIGH IN LIQ: 204.005

AUX: D:0.9971

DRY WEIGHT: 200.040

VOLUME: 3.59

Enter oil density

Enter the density of the oil being used to impregnate the part. Press Enter
button. Display advances to RETURN TO SETUP highlighted. Press Enter
button. Display returns to DENSITY SETUP with AUTO SAMPLE highlighted.

AUX LIQ & MODE

H20

<-SELECTED

OTHER

POROUS MATERIAL ON

DRY WEIGHT

200.040

DENSITY OF OIL

0.4000

RETURN TO SETUP

Testing the sample

Select RUN and press Enter button. Display advances to final DENSITY
display and requests to weigh the sample in air.

DENSITY SETUP

LIBRARY NAME

D2

SOLID DENSITY

<-SELECTED

LIQUID DENSITY

AUTO SAMPLE

OFF

AUTO PRINT

OFF

RUN

Now follow balance prompts to weigh sample in air and water. After weigh-
ing in air, weigh the sample in water. The balance will calculate Dry Density
and Oil Content by Volume (P1).

START/SAMPLE SETUP MAIN MENU

57.779

DENSITY - - D7

GRAMS/CC STABLE

Enter

DENSITY SETUP

Enter

LIBRARY NAME

D2

SOLID DENSITY

<-SELECTED

LIQUID DENSITY

AUTO SAMPLE

OFF

AUTO PRINT

OFF

RUN

Auto sample

Select AUTO SAMPLE and turn it ON or OFF. A setting of ON allows samples
to be sequentially sampled without pressing the Enter button for each
sample.

DENSITY SETUP

Enter

LIBRARY NAME

D2

SOLID DENSITY

<-SELECTED

LIQUID DENSITY

AUTO SAMPLE

OFF

AUTO PRINT

OFF

RUN

Auto print

At this point you can select automatic printing. To turn on, press the Enter
button and select ON. If you do not want automatic printing, select OFF.

P1: 535.58%

WEIGH IN AIR

To determine wet density

Wet density of the sample can be calculated by following the normal Solid
Density procedure using the oil impregnated part. When in AUX LIQ MODE
menu, turn OFF POROUS MATERIAL, then follow previous solid density
measuring procedure.

Appendix A - special applications