Final weighing of sieves, Appendix a - special applications – Ohaus Voyager Balances User Manual
Page 149

85
Enter
The above illustration represents 3 sieves which weighed between 3.4 and
3.9 grams each. TARE SETUP was set OFF, AUTO SAMPLE was set OFF and
% RETENTION was selected. The sample weight retained in each different
sieve weighed between 3.839 and 4.377 grams. The total amount of
retention is shown in percent on the last display which was 0.623 for three
sieves. Initial sample was 200.25 grams.
N/A
NA
3.426 3.839
N/A
N/A
3.969 4.388
N/A
N/A
3.962 4.377
CONTINUE SAVE RESAMPLE DELETE
TAREWT INIT WT TARE WT FINAL \VT
SAMPLE #1
With CONTINUE highlighted, press Enter button to review final display.
PRESS <ENTER> FOR MENU
SAMPLE # INIT WT FINAL WT % RETENSION
1
200.025 0.413 0.206
2
200.025 0.419 0.209
3
200.025 0.415 0.207
TOTAL
1.247 0.623
Final display indicates the initial weight, final weight of all sieves and the
total weight and total amount of retention in percent.
Final weighing of sieves
After pressing Enter button, the display returns to final weighing. Place first
full sieve on the pan and press Enter button. Repeat weighing with remain-
ing sieves pressing the Enter button after each sieve. When the last sieve
has been weighed, the display indicates the final weight of all samples.
3.841
DIFF WEIGHING
TARE WT:
TARE WT:
INIT WT: FINAL WT:3.841
PUT FULL SIEVE #1 ON PAN STABLE
START/SAMPLE EDIT SETUP
Appendix A - special applications