Set auto print on, Select number of samples, Enter sample size – Ohaus Voyager Balances User Manual

Page 123: Select pipette name, Enter pipette number, Enter pipette name, Select nominal, Select nominal value unit, Balance modes

Advertising
background image

59

Set auto print on

When set ON, results are automatically printed. Press the Enter button.
Display advances to PIPETTE SETUP with NUMBER OF SAMPLES highlighted.

Enter

Enter

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

Select number of samples

Press the Enter button. The display advances to NUMBER OF SAMPLES.

Enter sample size

Enter the number of samples. Press Enter button. The display advances to
PIPETTE SETUP with PIPETTE NAME highlighted.

AUTO PRINT

ON
OFF

10

1 10

NUMBER OF SAMPLES

Select pipette name

Press Enter button. Display advances to new screen PIPETTE NAME.

Enter pipette number

Enter the number of the pipette. Press the Enter button. The display ad-
vances to PIPETTE SETUP with NOMINAL highlighted.

Enter

PIPETTE SETUP

LIBRARY NAME

P2

TEST LIQUID(G/CC)

0.9971

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

PIPETTE NAME

Enter pipette name

Enter the name of the pipette. Press Enter button. Display advances to
PIPETTE NUMBER. Press Enter button.

WHEAT

PIPETTE NUMBER

345

Select nominal

Press the Enter button. The display advances to NOMINAL VALUE UNITS with
ML highlighted.

Enter

PIPETTE SETUP

BARO PRESSURE

1.0ATM

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

10

PIPETTE NAME

WHEAT

PIPETTE NUMBER

345

NOMINAL

0.00ML

Select nominal value unit

Select either ML (milliliters) or UL (microliters). Press the Enter button.
Display advances to NOMINAL.

NOMINAL VALUE UNITS

ML

UL

RETURN TO SETUP

or

Balance modes