Calibration test, 4 calibration test, 23 calibration – Ohaus Voyager Balances User Manual
Page 264: Enter place indicated mass value on pan, Enter, Clear the pan. enter
23
Calibration
4.4 Calibration test
Calibration test feature allows a check of a known calibration mass against the last stored calibration informa-
tion in the balance.
CALIBRATE
SPAN
LINEARITY
USER
CALIBRATION TEST
AUTOCAL
AUTOCAL DELTA CORRECTION
EXIT TO WEIGH
From main menu
Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to CALIBRATION TEST.
Enter
CALIBRATION TEST
PLEASE PUT 400.000 GRAMS
ON PAN AND PRESS <ENTER...
Enter
Place indicated mass value on pan.
CALIBRATION TEST
PLEASE WAIT. . .
THE DIFFERENCE BETWEEN THIS CAL
AND THE LAST CAL IS: 0.000 GRAMS
PRESS <ENTER> TO CONTINUE
Enter
Display indicates the difference between the present calibration and the
last calibration.
200.000
BASIC WEIGHING
Enter
0% 100%
Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.
GRAMS STABLE
PRESS <ENTER> FOR MENU
CALIBRATION TEST
PLEASE CLEAR THE PAN. . .
AND PRESS <ENTER>.
Clear the pan.
Enter