4.4 Calibration test

Calibration test feature allows a check of a known calibration mass against the last stored calibration information in
the balance.

CALIBRATE

SPAN

LINEARITY

USER

CALIBRATION TEST

AUTOCAL

AUTOCAL DELTA CORRECTION

EXIT TO WEIGH

From main menu

Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to CALIBRATION TEST.

Enter

CALIBRATION TEST

PLEASE PUT 400.000 GRAMS

ON PAN AND PRESS <ENTER...

Enter

Place indicated mass value on pan.

CALIBRATION TEST

PLEASE WAIT. . .

THE DIFFERENCE BETWEEN THIS CAL

AND THE LAST CAL IS: 0.000 GRAMS

PRESS <ENTER> TO CONTINUE

Enter

Display indicates the difference between the present calibration and the last
calibration.

200.000

BASIC WEIGHING

Enter

0% 100%

Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.

GRAMS STABLE

PRESS <ENTER> FOR MENU

CALIBRATION TEST

PLEASE CLEAR THE PAN. . .

AND PRESS <ENTER>.

Clear the pan.

Enter

Calibration