2 sensor administration under modbus, Sensor administration under modbus -6 – YSI DIQ/S 182 XT-4-MOD User Manual

Page 122

Advertising
background image

Modbus connection

System 182 XT-4

7 - 6

ba76028e01

02/2012

Query and response

telegram

Data transmission takes place via query and response telegrams. The
form of the telegrams is determined by the Modbus RTU protocol.
The Modbus master transmits a query telegram with a Modbus
command to the Modbus slave (here, the DIQ/S 182 XT-4-MOD). The
Modbus slave transmits a response telegram with the requested data
from the sensor data block or an error message.

Supported

Modbus commands

The DIQ/S 182 XT-4-MOD provides the data block for read access
only. Writing Modbus commands are not carried out.

7.4.2

Sensor administration under Modbus

Preparation of the IQ S

ENSOR

N

ET

for communication with the Modbus

may require, e.g. the following:

 Creating an IQ S

ENSOR

N

ET

system with specific assignment of

sensor numbers to sensors

 Creating several IQ S

ENSOR

N

ET

systems with the identical

assignment of sensor numbers to sensors

 Changing the order of the sensors in an already installed system

Creating the assignment

of sensor numbers

You want to install an IQ S

ENSOR

N

ET

system and, at the same time,

create a specific sequence of sensor number assignments to the
sensors.

Modbus commands

Command code

Read Input Register

04h

Read Discrete Inputs

02h

Read Holding Register

03h

Read Coils

01h

1

Carry out a system start without any sensors (see section
3.12).

2

Connect the sensors to the system in the required order. With
each sensor wait until the system has recognized the sensor.

Advertising