11 connection to the relay outputs, 1 general installation instructions, 11 connection to the relay outputs -42 – YSI DIQ/S 182 XT-4-MOD User Manual

Page 56: 1 general installation instructions -42

Advertising
background image

Installation

System 182 XT-4

3 - 42

ba76028e01

02/2012

3.11

Connection to the relay outputs

3.11.1 General installation instructions

Warning

If external electrical circuits that are subject to the danger of
physical contact are incorrectly connected to the relay contacts,
there may be a danger of life threatening electric shock. Electrical
circuits are regarded to be subject to the danger of physical
contact when there are voltages higher than the Safety Extra Low
Voltage (SELV).

Pay attention to the following points during installation:
Electrical circuits subject to the danger of physical contact

must only be connected by a qualified electrician.

Electrical circuits subject to the danger of physical contact

must only be connected when they are voltage-free.

If electrical circuits subject to the danger of physical contact

are switched with a relay, no circuit that is not subject to this
danger (e. g. the DIQ/CHV module) may be operated on the
further relays.

Switching voltages and switching currents on the relay

contacts must not exceed the values specified in
chapter 10 T

ECHNICAL

DATA

. Protect electrical circuits against

currents that are too high with an electrical fuse.

Only single-phase consumers can be switched with the relays.

Under no circumstances must multiphase consumers be
switched with the aid of several relays (example three-phase
current driven pumps). Always switch multiphase consumers
via a protective relay.

The DIQ/S 182 XT-4-MOD may only be opened after the

installation if all external voltages have been previously
switched off.

Connections

The DIQ/S 182 XT-4-MOD has three relay outputs in the front part of
the enclosure. The terminal strips are shown in section 3.14.

Materials required

 Wire end sleeves, suitable for the connecting wires, with suitable

crimping tool

 1 x cable gland, matched to the cable diameter (see section 3.9 on

page 3-33).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screwdriver

 Small screwdriver