System 182 XT-4

Safety instructions

2 - 3

ba76028e01

02/2012

Safe operation

If safe operation is no longer possible, the System 182 XT-4 must be
taken out of operation and secured against inadvertent operation.
Safe operation is no longer possible if components:

 have been damaged in transport

 have been stored under adverse conditions for a prolonged period

of time

 are visibly damaged

 no longer operate as described in this manual.

If you are in any doubt, contact the supplier of your System 182 XT-4.

Obligations of the

operator

The operator of the System 182 XT-4 must ensure that the regulations
and guidelines listed below are followed when dealing with dangerous
substances:

 EEC guidelines relating to safety at work

 National laws relating to safety at work

 Accident prevention regulations

 Safety datasheets of the chemicals manufacturers.

Warning

A circuit (except for the power supply connections) that is
connected to the System 182 XT-4 must not feed any voltages or
currents that are not allowed. It has to be made sure that the
circuit at any time meets all requirements of a Limited circuit
or
Limited Power
as well as of SELV (Safety Extra Low Voltage).
These include the following limiting value specifications:
AC voltage: max. 30 V effective / 42.4 V peak
DC voltage: max. 60 V
Current limit: max. 8 A
Power output limitation: max. 150 VA