3 sensor data block, Sensor data block -8 – YSI DIQ/S 182 XT-4-MOD User Manual

Page 124

Advertising
background image

Modbus connection

System 182 XT-4

7 - 8

ba76028e01

02/2012

7.4.3

Sensor data block

The data block contains the data of the sensors in the order of their
registering on the DIQ/S 182 XT-4-MOD (Sxx number).

The data of a sensor are stored in 8 registers each.
Thus, the sensor data block consists of altogether 16 registers. Each
register contains 2 bytes. The bytes 0 and 1 form register 0001, the
bytes 2 and 3 form register 0002, etc.

Modbus

Information

Data format

Register Bit

Data type

Bit

Sensor S01

0001

1-8

1 [= sensor number (S01)]

(Int 8)

7-0

0001

9-16

Sensor status

(Int 8)

7-0

0002

1-16

Sensor model

(Int 16)

15-0

0003

1-16

Status info

(Int 16)

15-0

0004

1-8

Measuring mode

(Int 8)

7-0

0004

9-16

Measured value status

(Int 8)

7-0

0005

1-16

Main measured value (IEEE-754 floating point, 32-bit)

(Float 32)

31-16

0006

1-16

Main measured value (IEEE-754 floating point, 32-bit)

(Float 32)

15-0

0007

1-16

Secondary meas. value (IEEE-754 floating point, 32-bit)

(Float 32)

31-16

0008

1-16

Secondary meas. value (IEEE-754 floating point, 32-bit)

(Float 32)

15-0

Sensor S02

0009

1-8

2 [= sensor number (S02)]

(Int 8)

7-0

0009

9-16

Sensor status

(Int 8)

7-0

0010

1-16

Sensor model

(Int 16)

15-0

0011

1-16

Status info

(Int 16)

15-0

0012

1-8

Measuring mode

(Int 8)

7-0

0012

9-16

Measured value status

(Int 8)

7-0

0013

1-16

Main measured value (IEEE-754 floating point, 32-bit)

(Float 32)

31-16

0014

1-16

Main measured value (IEEE-754 floating point, 32-bit)

(Float 32)

15-0

0015

1-16

Secondary meas. value (IEEE-754 floating point, 32-bit)

(Float 32)

31-16

0016

1-16

Secondary meas. value (IEEE-754 floating point, 32-bit)

(Float 32)

15-0

Advertising