9.1

Information on errors

Log book

The 182 XT-4 system performs a comprehensive cyclical self test
during operation. While doing so, the system identifies all states that
deviate from normal operation and enters corresponding messages in
the log book (information or error message).

With the aid of the log book, you can call up instructions on how to clear
the error directly on the terminal. The log book is described in detail in
section 4.4.3 S

ENSOR

STATUS

S

XX

.

Note
Information on possible errors in IQ sensors and MIQ output modules
is given in the chapter W

HAT

TO

DO

IF

... of the respective component

operating manual.

9.2

Error causes and remedies

Note
Errors of the Modbus communication are dealt with in detail in
section 7.6.

The system does no

longer react on entries

"Error" in the measured

value display

Cause

Remedy

– System error

Reset the system:
– Switch off the power supply

and switch it on again after
10 s

Cause

Remedy

– Communication with the

IQ Sensor interrupted

– Check the cable connection

– Error in the IQ sensor

– Disconnect the IQ sensor

and reconnect it after 10 s