YSI DIQ/S 182 XT-4-MOD User Manual
Page 57
System 182 XT-4
Installation
3 - 43
ba76028e01
02/2012
Connecting lines to the
terminal strip
Fig. 3-31 Feeding in the lines (example: front terminal strip)
Warning
No free wires must be allowed to project into the enclosure.
Otherwise, there is a danger that areas safe to contact could come
into contact with dangerous voltages. This could result in life
threatening electric shock when working with the
System 182 XT-4. Always cut off any wires that are not in use as
closely as possible to the cable gland.
11
Open the enclosure of the DIQ/S 182 XT-4-MOD.
12
Screw a cable gland (pos. 1 in Fig. 3-31) with the sealing ring
(pos. 2) into the enclosure below the respective connections.
13
Loosen the coupling ring (pos. 3).
14
Feed the line through the cable gland in the enclosure.
15
Connect the wires to the terminal strip. While doing so, pay
attention to the specifications on the label located under the
terminal strip.
16
Tighten the coupling ring (pos. 3).
1
3
2