System 182 XT-4

Installation

3 - 39

ba76028e01

02/2012

3.10.2 DIQ/S 182 XT-4-MOD/24V (24 V version)

Warning

If the 24 V AC/DC supply is incorrectly connected, it may
represent a danger to life from electric shock. Pay attention to the
following points during installation:
The DIQ/S 182 XT-4-MOD must only be connected by a

qualified electrician.

The 24 V AC/DC supply must meet the specifications quoted on

the name plate and in chapter 10 T

ECHNICAL

DATA

(protective

low voltage SELV).

The DIQ/S 182 XT-4-MOD must only be connected in a voltage-

free condition.

A switch or circuit breaker must be provided in the building

installation as an interrupt facility for the DIQ/S 182 XT-4-MOD.
The interrupt facility must be

– installed in the vicinity of the DIQ/S 182 XT-4-MOD, easily

accessible by the user and

– marked as an interrupt facility for the DIQ/S 182 XT-4-MOD.

Note
Rechargeable battery systems should have a deep discharge
protection. The DIQ/S 182 XT-4-MOD/24V does not have any built-in
deep discharge protection.

Materials required

 Wire end sleeves, suitable for the 24 V AC/DC feed line, with

suitable crimping tool

 1 x cable gland, matched to the cable diameter (see section 3.9 on

page 3-33).

Tools

 Cable stripping knife

 Wire stripper

 Phillips screwdriver

 Small screwdriver.

Preparing the 24 V AC/

DC line

1

Cut off the cable to the required length.

2

Strip the cable insulation for approx. 45 mm.

3

Bare the wires 1 and 2 and fit them with wire end sleeves.