Fieldbus

11.1 Modbus register information

Table 8 Sensor modbus registers

Tag Name

Register #

Data
Type

Length R/W

Discrete
Range

Min/Max
Range

Description

MEASURE VALUE 1 40001

Float

2

R

actual measurement value from
channel one

LOCATION 1

40005

String

8

R/W

Name of LOCATION 1 (see menu
system)

MEAS.

40013

Float

2

R


0/99999.9

measuring value ; photometer data

MEAS. ZERO

40015

Float

2

R


0/99999.9

measuring zero; photometer data

OFFSET MEAS.

40017

Float

2

R


0/99999.9

offset measuring; photometer data

REF

40019

Float

2

R


0/99999.9

reference value; photometer data

REF ZERO

40021

Float

2

R


0/99999.9

reference zero; photometer data

OFFSET REF

40023

Float

2

R


0/99999.9

offset reference; photometer data

AMPLIFY MEAS.

40025

Integer

1

R

-100/100

amplification of measurement
channel

AMPLIFY REF

40026

Integer

1

R

-100/100

amplification of reference channel

COOLING

40027

Unsigned
Integer

1

R

0/100

percentage of cooling fan power

HEATING

40028

Unsigned
Integer

1

R

0/1

status off heating for sample tube;
0=OFF, 1=ON

HUMIDITY PROBE

40029

Unsigned
Integer

1

R

0/100

humitity filtration probe in percent

ENCLOSURE
TEMP

40030

Float

2

R

0/99.9

the temperature inside the analyzer

ACTUAL VALUE

40032

Float

2

R

actual measurement value from
channel one

LOCATION 2

40034

String

8

R/W

location for the measuring channel
two where the sample is coming
from

CONTENT

40042

Unsigned
Integer

1

R

0/65535

the entry is for the device driver
file;shows the version

REMAINING TIME

40043

Unsigned
Integer

1

R

0/65535

remaining time of the current
process

CUVETTE TEMP.

40044

Float

2

R

0/99.99

actual cuvette temperature

PO4-P VALUE 1

40046

Float

2

R

measurement value for channel one
as PO4-P

PO4-P VALUE 2

40048

Float

2

R

measurement value for channel two
as PO4-P

PO4 VALUE 1

40050

Float

2

R

measurement value for channel one
as PO4

PO4 VALUE 2

40052

Float

2

R

measurement value for channel two
as PO4

P2O5 VALUE 1

40054

Float

2

R

measurement value for channel one
as P2O5