Troubleshooting, 3 warnings (continued) – Hach-Lange PHOSPHAX indoor sc User Manual

Page 50

Advertising
background image

50

Troubleshooting

PHOT LEVEL LOW /
PHOT LEVEL2 LOW

Continued
measurement

Photometer signal low
(dependent on chanel if 2
chanel mode is used) .
Not enough sample.

Trigger cleaning (several
times). Manually clean
photometer. If this action
solves the problem,
increase automatic
cleaning, otherwise contact
service. Check sample
delivery.

Automatic

MODULES CONTAM.

Continued
measurement

Filter modules soiled

Clean filter modules soon

Automatic

SERVICE MODE

Service state

The instrument is in the
service state or is
switching to this state

Automatic when
the service state
is left

REAGENT LEVEL

Continued
measurement

Amount of reagent has
dropped below warning
level set

Check reagent level and
replace if necessary, then
reset reagent level. The
level is indicated
mathematically and can only
function reliably if the
counter is only reset when
the solution is changed

On the menu
MAINTENANCE/
TEST/MAINT
COUNTER/
REAGENT

CLEAN SOLU LEVEL

Continued
measurement

Amount of cleaning
solution has dropped
below warning level set

Check cleaning solution
level and replace if
necessary, then reset
cleaning solution level.
The level is indicated
mathematically and can only
function reliably if the
counter is only reset when
the solution is changed

On the menu
MAINTENANCE/
TEST/MAINT
COUNTER/
CLEANING
SOLU.

7.2.3 Warnings (continued)

Warning displayed

Instrument
reaction

Cause

Solution

Reset warning

Advertising