Installation

3.6

Installation of reagents

CAUTION
Good safety habits and technique should be observed at all
times. Review all SDS/MSDS information and use
recommended safety measures to prevent exposure to
potential chemical hazards.

CAUTION
Avoid unnecessary contact with sample flows of unknown
concentration as they can produce hazards due to trace
chemicals, radiation, or biological effects.

CAUTION
The enclosure may tip forwards if it has not been fixed in
place. Only open the enclosure if the enclosure is properly
mounted.

Important Note: Always lay the drain tubing so that there is a
continuous fall (minimum 3°), the outlet is clear (not pressurized)
and, the drain tubing is not longer than 2 meters. For more
information refer to

Appendix A on page 65

.

Important Note: Incorrect use of the reagents can damage the
instrument. Carefully read the labels on the containers to ensure
there are no mistakes.

1. Place the reagent containers in the instrument (

Figure 17

).

2. Insert the tubing in the reagent containers.

Note: If the tubing connections are twisted, no chemicals will flow
through the connection and the instrument will not work correctly. Turn
the bottle while holding the lid still so that the tubing connections are
not twisted.

3. Screw the reagents to the caps supplied.