Plumbing and connection options

A.10 Option 6 plumbing and connections

Option 6 uses two sc analyzers with two FILTRAX (FILTRAX 1 and
FILTRAX 2). Samples from both FILTRAX are going into Analyzer 1
using the 2-parameter configuration. The heated drain hose
connects both sc analyzers. The waste from both analyzers is
discharged into a drain through the heated drain hose.

Refer to

Figure 32

and the following instructions for Option 6:

1. Install both FILTRAX into the sample stream. Refer to the

FILTRAX User Manual for more information.

2. Install the first sc analyzer (Analyzer 1):

a. Feed the heated hose from FILTRAX 1 through the

analyzer (

Figure 32

, item 25). Use Seal Plug #1 to secure.

b. Feed the heated hose from FILTRAX 2 through the

analyzer (item 27). Use Seal Plug #1 to secure.

c. Feed the heated drain hose through the analyzer (item 28).

Use Seal Plug #1 to secure. Connect the heated drain
power connections. Refer to

section 3.5.5, page 27

.

d. Remove the pre-installed drain tube attached to the valve

block and remove the T-fitting from the drain tube. Retain
for reuse with Analyzer 2.

e. Connect the heated drain tube to the valve block connector.

f.

Connect the sample line from FILTRAX 1 to the bottom inlet
on Overflow Vessel 1 using the fittings (item 20).

g. Connect the sample line from FILTRAX 2 to the bottom inlet

on Overflow Vessel 2 using the fittings (item 19).

h. Change the analyzer to the 2-parameter configuration.

Refer to

Connect a 2-parameter option, page 66

.

i.

Connect Sample Line 1 from the heated drain to Overflow
Vessel 1. Connect Sample Line 2 from the heated drain to
Overflow Vessel 2.

3. Install the second sc analyzer (Analyzer 2)

a. Feed the heated drain hose from Analyzer 1 through

Analyzer 2 (item 13). Use Seal Plug #1 to secure.

b. Feed the heated drain hose through the analyzer (item 8).

Use Seal Plug #1 to secure. Connect the heated drain
power connections. Refer to

section 3.5.5, page 27

.

c. Seal the remaining opening with Seal Plug #3.

d. Cut 25 mm from the drain tube that was removed from

Analyzer 1. Connect the 25 mm piece of tubing to the
T-fitting on Analyzer 2. Connect the other end of the tubing
to the T-fitting that was removed from Analyzer 1. Refer to

Figure 26, page 67

for T-fitting removal.

e. Connect the drain tube from Analyzer 1 and the drain tube

from Analyzer 2 to the T-fitting.

4. Connect Sample Line 1 from Analyzer 1 to Overflow Vessel 1

using the fittings (item 16). Connect Sample Line 2 from
Analyzer 1 to Overflow Vessel 2 using the fittings (item 17).