Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual

Page 102

Advertising
background image

102

Fieldbus

P2O5 VALUE 2

40056

Float

2

R

measurement value for channel two
as P2O5

DEXT LAST VALUE

40058

Float

2

R


-9.99999/
9.99999

last extinction value

HEATING ON

40060

Unsigned
Integer

1

R/W


0/1/2/3/4/
5/6/7/8/9/
10/11/12

sets the month when the sample
line heating is switched ON;
0=alway OFF, 1=January,
2=Febraury to 12=December

HEATING OFF

40061

Unsigned
Integer

1

R/W


1/2/3/4/5/
6/7/8/9/1
0/11/12

sets the month when the sample
line heating is switched
OFF;1=January, 2=Febraury to
12=December

MEASURE VALUE 2 40062

Float

2

R

actual measurement value from
channel two

EXT MESS 1

40064

Float

2

R


-9.99999/
9.99999

the current measuring extinction
from channel one

EXT MESS 2

40066

Float

2

R


-9.99999/
9.99999

the current measuring extinction
from channel two

EXT REF

40068

Float

2

R


-9.99999/
9.99999

the current reference extinction

ANALYZER
HEATING

40070

Unsigned
Integer

1

R

0/100

the heating of the analyzer

BUSACTION
ACTIVE

40071

Unsigned
Integer

1

R/W

0/1

write a one to this register to start a
bus action (see Field bus control)

BUS ANALY.START

40072

Unsigned
Integer

1

R/W

0/1

you can start the analyzer about the
bus

BUS SERVICE

40073

Unsigned
Integer

1

R/W

0/1

you can start the service mode
about the bus

BUS CLEANING

40074

Unsigned
Integer

1

R/W

0/1

you can start the cleaning mode
about the bus

BUS PREPUMP
REA.

40075

Unsigned
Integer

1

R/W

0/1

you can prepump reagent about the
bus

BUS PREPUMP
CLEA

40076

Unsigned
Integer

1

R/W

0/1

you can prepump cleaning solution
about the bus

BUS PREPUMP
PRO.

40077

Unsigned
Integer

1

R/W

0/1

you can prepump the probe about
the bus

BUS PREPUMP
ALL

40078

Unsigned
Integer

1

R/W

0/1

you can prepump all about the bus

PROBE P. MIN

40079

Float

2

R

0/2.0

integrated value of pressure at
filtration probe, if not yet calculated:
nan

PROBE
PRESSURE

40081

Float

2

R

0/2.0

actual value of pressure at filtration
probe, if not yet calculated: nan

GAIN CORR. 1

40083

Float

2

R/W

0.01/
100.00

Gain correction for channel one

DISCHARGE
CLEAN.

40096

Unsigned
Integer

1

R/W

0/10

discharged values after a cleaning

Table 8 Sensor modbus registers