Hach-Lange PHOSPHAX indoor sc User Manual

Page 5

Advertising
background image

5

Section 1

Specifications

Specifications are subject to change without notice.

Enclosure rating

PHOSPHAX sc: IP55
PHOSPHAX indoor sc: IP54

Enclosure material

ASA/PC UV-resistant

Measuring method

2-beam photometer (yellow method)

Measuring range

0.05 to 15 mg/L PO

4

–P

1 to 50 mg/L PO

4

–P

Detection limit

0.05 mg/L, with standard solution: (0.05 to 15 mg/L PO

4

–P Measuring Range)

1.00 mg/L, with standard solution (1 to 50 mg/L PO

4

–P Measuring Range)

Measuring accuracy
(with standard solution)

2 % of the measured value + 0.05 mg/L (0.05 to 15 mg/L PO

4

–P Measuring Range)

2 % of the measured value + 1.0 mg/L (1 to 50 mg/L PO

4

–P Measuring Range)

Repeatability
(with standard solution)

2 % of the measured value + 0.05 mg/L (0.05 to 15 mg/L PO

4

–P Measuring Range)

2 % of the measured value + 1.0 mg/L (1 to 50 mg/L PO

4

–P Measuring Range)

Response time (90 %)

< 5 minutes

Adjustable measuring interval

5 to 120 minutes

Power supply

Power supply with power cable, using the sc1000 controller only
(analyzer, Filter Probe sc, and drain tubing: 115 V versions or 230 V versions)

Data transmission

Data transmission with data cable on the sc1000 controller

Electrical power consumption

500 VA

Electrical fuse protection

Via sc1000 controller
Maximum of 2 analysis instruments for each sc1000.

Outputs

Relay, current outputs, network interface via sc1000 controller. Refer to the sc1000
manual for detailed specifications for analog, relay, and digital outputs.

Operating temperature

PHOSPHAX sc: –20 to 45 °C (–4 to 113 °F); 95 % relative humidity,
non-condensing
PHOSPHAX indoor sc: 5 to 40 °C (41 to 104 °F); 95 % relative humidity,
non-condensing

Storage temperature

–20 to 60 °C (–4 to 140 °F); 95 % relative humidity, non-condensing

Sample temperature

+4 to +45 °C (39 to 113 °F)

Sample pressure

With continuous sample preparation –30 mbar to +50 mbar at overflow vessel

Sample flow

Range: 1.0–20.0 L/h

Sample quality

Ultra filtrated or

comparable

Sample level

Level of liquid in basin with filtration probe must be below analyzer

Permissible pH value of the sample

5 to 9

Permissible chlorid range

 1000 mg/L Cl

Dimensions (

Figure 1 on page 6

,

Figure 2 on page 7

)

PHOSPHAX sc: (W × H × D) 540 Ч 720 Ч 390 mm (21.25 Ч 28.35 Ч 15.35 in.)
PHOSPHAX indoor sc: (W × H × D) 540 Ч 720 Ч 370 mm (21,25 Ч 28,35 Ч 14.5 in.)

Data and power cable lengths

2 m (80 in.) (from edge of enclosure)

Weight

PHOSPHAX sc: Approximately 31 kg, without Filter Probe sc and without
chemicals
PHOSPHAX indoor sc: Approximately 29 kg, without Filter Probe sc and without
chemicals

Advertising