2 sensor setup menu (continued), Operation – Hach-Lange PHOSPHAX indoor sc User Manual

Page 38

Advertising
background image

38

Operation

MAINTENANCE (continued)

MAINT. COUNTER

Counter for reagent and consumables

OPERATING HOURS

Displays the instrument's operating hours.

REAGENT

Displays the current level of the reagent.

CLEANING SOL.

Displays the current level of the cleaning solution.

AIR FILTER PADS

Days left until the next air filter change/clean.

PISTON PUMP

Days left until the next pump piston and cylinder replacement (PHOSPHAX piston pump)

REAGENT PUMP

Number of pump strokes performed by the reagent metering pump.

STATUS MODULES

Displays the state of the modules (on filtration probe operation).

CLEANING MODULES

Last filter module cleaning (on filtration probe operation).

NEW MODULES

Last filter module replacement (on filtration probe operation).

PUMP MEMBRANE

Date of the last pump membrane replacement (filtration probe sample pump) (on filtration
probe operation).

COMPRESSOR

Days left until the replacement of the air compressor (on filtration probe operation).

TEST/MAINT

Maintenance processes

SIGNALS

PROCESS

Indication of what the instrument is doing.

REMAINING TIME

Indication of the time left for the currently ongoing process

CUVETTE TEMP.

Current measuring cell temperature

ENCLOSURE TEMP

Current temperature in the instrument

COOLING

Current speed of the enclosure fan in %

HEATING

Current enclosure heating power

PRESSURE ANALY

Current pressure in the metering system of the valve block in mbar

HUMIDITY ANALY

Indication of whether there is liquid in the collecting tray

STATUS MODULES

Only if filtration probe is registered: shows status of filtration modules (0–100 %)

PRESSURE P. MIN

Only if filtration probe is registered :shows the averaged minimum pressure at the filter
modules

PRESSURE PROBE

Only if filtration probe is registered: shows the actual minimum pressure at the filter modules

PROBE HEATING

Only if filtration probe is registered: indication of switching state of sample tubing heating

HUMIDITY PROBE

Only if filtration probe is registered: indication of whether there is moisture in the probe
enclosure

DRAIN HEATING

only in 1 or 2 channel mode: shows status of drain heating

PROCESS

Indication of what the instrument is doing.

REMAINING TIME

Indication of the time left for the currently ongoing process

SERVICE MODE

Instrument can be set to service mode, for example for maintenance (system free of liquids,
thermal management and compressor for filter probe (if installed) active)

SET OUTMODE

Value that is output in the service state. HOLD=last measured value, SET TRANSFER= value
to be entered

START

Leave service mode, start measurement

REAGENT

Reset the maintenance counter after reagent change.

CLEANING SOL.

Reset the maintenance counter after changing the cleaning solution.

AIR FILTER PADS

Menu-based process for changing the air filter pads, resetting the maintenance counter.

PISTON PUMP

Days left until the next pump piston and cylinder replacement (PHOSPHAX piston pump),
reset after replacement of the pump

REAGENT PUMP

Number of pump strokes performed by the reagent metering pump, reset after replacement of
the pump.

5.2 Sensor setup menu (continued)