46

Maintenance

2. On the controller TEST/MAINT menu, start a cleaning cycle

with distilled water using the FLUSHING.

3. Clean the canister lid with distilled water.

4. Take the tubing out of the water and start the FLUSHING

function to pump the tubing and the analysis instrument empty.

5. Wipe the canister lids dry and seal the canisters with the

corresponding lids.

6. Remove the canisters and store them in a frost-free place and

in accordance with local regulations.

7. Isolate the system from the mains and the data network.

8. When using a Filter Probe sc, refer to the Filter Probe sc User

Manual for storage information.

9. Install all transport locks.

10. Depending on the duration, remove the system from its

mounting and wrap the system in a protective film or dry cloth.
Store the system in a dry place.

6.5 Scheduled maintenance

Table 5

lists items that need to be maintained by service personnel

ONLY. Contact the manufacturer for more information.

6.6 Modify from single channel to dual channel

The sc analyzer can be converted from single channel to dual
channel operation and/or continuous sampling. Contact the
manufacturer for more information. Refer to

Table 6

for

configuration options.

Table 5 Repair maintenance items

Description

When to replace

Warranty

Reagent pump for sc analyzer (Valve pumps)

According to wear

1 year

Pump head piston pump 10 mL
(Pre-greased cylinder and piston)

1 year

1 year

Switchable Compressor 115/230V

2 years recommended

2 years

Table 6 Conversions

From

To

With

Conversion kit

1-channel operation

2-channel operation

AMTAX sc, PHOSPHAX sc

LZY170

Filter Probe sc

Continuous sampling

AMTAX sc, PHOSPHAX sc

LZY241

Continuous sampling

Filter Probe sc

AMTAX sc, PHOSPHAX sc

LZY242