Plumbing and connection options

3. Install the second sc analyzer (Analyzer 2):

a. Feed the heated drain hose from Analyzer 1 through

Analyzer 2 (item 12). Use Seal Plug #1 to secure.

b. Feed the second heated drain hose through Analyzer 2

(item 11). Use Seal Plug #1 to secure.

c. Seal the remaining opening with Seal Plug #3.

d. Connect the heated drain power connections. Refer to

section 3.5.5, page 27

.

e. Cut 25 mm from the drain tube that was removed from

Analyzer 1. Connect the 25 mm piece of tubing to the
T-fitting on Analyzer 2. Connect the other end of the tubing
to the T-fitting that was removed from Analyzer 1. Refer to

Figure 26, page 67

for T-fitting removal.

f.

Connect the drain tube from Analyzer 1 and the drain tube
from Analyzer 2 to the T-fitting.

4. Connect the sample line from Analyzer 1 to the bottom inlet on

the overflow vessel.