Plumbing and connection options

A.16 Option 10a plumbing and connections

Option 10a uses an sc analyzer with any type of sample
preparation that delivers a continuous sample stream that cannot
be pressurized. The waste of the analyzer is discharged into an
open drain.

Refer to

Figure 38

and the following instructions for Option 10a:

1. Install the sample preparation unit.

2. Feed the sample line from the sample preparation unit through

the analyzer (

Figure 38

, item 2). Use Seal Plug #3 to secure.

3. Feed the drain tube through the analyzer (item 3). Use Seal

Plug #3 to secure.

Note: Tubes can be pushed through prepared holes on Seal Plug #3.

4. Seal any unused holes with Seal Plug #3.

5. Connect the drain tube to the T-fitting.

6. Connect the sample line from the sample preparation to the

bottom inlet on the overflow vessel using the fittings.