3 warnings, Troubleshooting, 2 error messages (continued) – Hach-Lange PHOSPHAX indoor sc User Manual

Page 49

Advertising
background image

49

Troubleshooting

PHOTO LEVEL
HIGH / PHOTO
LEVEL2 HIGH

Continued
measurement

Photometer signal
too high (1
channel, 2 channel
operation ,
depending on
channel that
triggered the error)

Contact service

Reset error
manually or
automatic if level is
OK again

MODULES
CONTAM.

Continued
measurement

Filter modules
heavily soiled

Clean filter modules immediately

Reset error
manually

DRAIN BLOCKED

Service state

Drain is blocked

Clean drain line

Reset error
manually

SAMPLE1 /
SAMPLE2

Continued
measurement

the amount of
sample is not
sufficient
(channel1 /
channel2) This
occurs as error if
SAMPLE
DETECTION is set
to ERROR

Check sample delivery, make sure that
sample line has no negative pressure,
check tightness of piston pump, check
overflow and air valve

Automatic reset
when enough
sample is
available, or
manuell reset

7.2.2 Error messages (continued)

Error displayed

Instrument
reaction

Cause

Solution

Reset error

7.2.3 Warnings

Warning displayed

Instrument
reaction

Cause

Solution

Reset warning

WARMUP PHASE

Instrument is
warming up sample
tubing after start
(de-frosting)

If there is a risk that the
sample tubing is frozen, a
warning is displayed

As far as possible, wait until
end of the warm-up phase
(except if certain there is no
frost), to cancel please
place instrument in service
state and start
measurement again

Automatic

COOLING DOWN

Fan 100 % ,
standstill until cool
enough

Instrument is cooling down
after start using ventilation,
if it was heated
excessively

Wait until instrument has
cooled down enough

Reset,
automatically as
soon as cool

ANALYZER TO COLD

Measurement

Instrument has cooled
below 15°C (59 °F)

Close instrument door, if
necessary check heating

Reset,
automatically as
soon as warmer

ANALYZER TO WARM

Measurement, but
no more air
cleaning

At very high internal
temperature, the air
cleaning of the filter
modules is disabled to
generate less wasteheat of
the internal compressor.
(internal temp. =
55°C(131°F ))

Change/clean air filter,
check air ducts for blockage,
check enclosure fan, is
ambient temperature
allowed?

Reset,
automatically as
soon as cool

CUV TOO COOL

Continued
measurement

The cuvette is not being
heated adequately. 2 min
after sample change:
temp. = ((target temp. of
cuvette) - 1 °C(34°F))

Close instrument door,
check/fit cuvette insulation.

Automatic