Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual

Page 105

Advertising
background image

105

Fieldbus

CLEAN SOLU
LEVEL

40158

Unsigned
Integer

1

R

0/100

cleaninig solution level in percent

AIR FILTER DISPL

40160

Integer

1

R

-32768/
32767

days until next cleaning/exchange
of air filter pads. Negative values
show a cleaning which is overdue

PUMP DISPLAY

40163

Integer

1

R

-32768/
32767

days left until exchanging piston of
pump, negative values show that
exchange is overdue

STRUCTURE

40172

Unsigned
Integer

1

R

0/65535

the entry is for the device driver
file;shows the version

FIRMWARE

40173

Unsigned
Integer

1

R

0/65535

the entry is for the device driver
file;shows the version

LOADER

40174

Float

2

R


0/3.40282
34663852
9E+38

the entry is for the application
file;shows the version of the boot file

OPERATING
HOURS

40176

Unsigned
Integer

2

R


0/999999
99

operating hours of analyzer

PUMP
MEMBR.DISP.

40180

Integer

1

R

-32768/
32767

remaining days for pump membrane
in filter probe

COMPRESSOR

40189

Integer

1

R

-32768/
32767

remaining days for air compressor

LAST
CHA.FACTOR1

40198

Time2

2

R

the date of the last correction factor
for channel one

LAST
CHA.FACTOR2

40200

Time2

2

R

the date of the last correction factor
for channel two

SAMPLE
DETECTION

40211

Unsigned
Integer

1

R/W

0/1/2

output if sample detection detects
low sample amount; 0=Warning,
1=Error, 2=OFF

ACTUAL
MEAS.TIME

40216

Time2

2

R

time of actual measurement value

LAST TIME

40218

Time2

2

R

time of last measurement value

2.ND LAST TIME

40220

Time2

2

R

2.ND LAST TIME

3.RD LAST TIME

40222

Time2

2

R

3.RD LAST TIME

4.TH LAST TIME

40224

Time2

2

R

4.TH LAST TIME

5.TH LAST TIME

40226

Time2

2

R

5.TH LAST TIME

6.TH LAST TIME

40228

Time2

2

R

6.TH LAST TIME

7.TH LAST TIME

40230

Time2

2

R

7.TH LAST TIME

8.TH LAST TIME

40232

Time2

2

R

8.TH LAST TIME

9.TH LAST TIME

40234

Time2

2

R

9.TH LAST TIME

LAST VALUE

40236

Float

2

R

LIST OF VALUES

2.ND LAST VALUE

40238

Float

2

R

LIST OF VALUES

3.RD LAST VALUE

40240

Float

2

R

LIST OF VALUES

4.TH LAST VALUE

40242

Float

2

R

LIST OF VALUES

5.TH LAST VALUE

40244

Float

2

R

LIST OF VALUES

6.TH LAST VALUE

40246

Float

2

R

LIST OF VALUES

7.TH LAST VALUE

40248

Float

2

R

LIST OF VALUES

Table 8 Sensor modbus registers

Tag Name

Register #

Data
Type

Length R/W

Discrete
Range

Min/Max
Range

Description