Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual

Page 104

Advertising
background image

104

Fieldbus

HUMIDITY ANALY

40127

Unsigned
Integer

1

R

0/100

humidity analyzer in percent

FIELDBUS

40128

Unsigned
Integer

1

R/W

0/2

triggers a measurement serie in
START BY BUS mode (see Start by
bus)

START BY BUS

40129

Unsigned
Integer

1

R/W

0/1

sets the analyzer to START BY BUS
mode ( see Start by bus)

EXTINCTION 2

40130

Float

2

R


-9.99999/
9.99999

the curent extinction from the
measurement channel two

PROCESS STATE

40132

Unsigned
Integer

1

R


0/1/2/3/4/
5/6/7/8/9/
10/11/12/
13/14/15/
16

the process/state of the analyzer,
coded as enum list; enum value
0=service mode, enum value
1=measuring 1..., interval,
initialisation, serv.in proc., cleaning,
warm up phase, measuring 2...,
prepump reag., prepump clean.,
prepump probe, flushing, start by
bus, calibrate, test procedure,
prepump.sample, enum value
16=validation

NUMBER OF
MEAS.

40133

Unsigned
Integer

1

R/W

1/100

Number of measurements in a
START BY BUS measurement
series (see Start by bus)

AVERAGE

40134

Unsigned
Integer

1

R/W

?

The number of measurement values
that result in an average value in a
START BY BUS measurement
series.

NO.OF VALUES
CH1

40135

Unsigned
Integer

1

R/W

0/100

2 channel mode: how often is
channel 1 measured before
switching to channel 2

NO.OF VALUES
CH2

40136

Unsigned
Integer

1

R/W

0/100

2 channel mode: how often is
channel 2 measured before
switching to channel 1

DISCHARGE VAL1

40137

Unsigned
Integer

1

R/W

0/3

number of discharged values when
switching from channel 1 to channel
2

DISCHARGE VAL2

40138

Unsigned
Integer

1

R/W

0/3

number of discharged values when
switching from channel 2 to channel
1

DISCHARGE BUS

40140

Unsigned
Integer

1

R/W

2/10

discharge values at the beginning of
a START BY BUS series

SENSOR NAME

40143

String

8

R

User-assigned name for a sensor

REAG. WARNING

40151

Unsigned
Integer

1

R/W

0/1

warning if the level of reagent is low;
0=OFF, 1=ON

WARNING

40152

Unsigned
Integer

1

R/W


20/15/10/
5

Reagent warning level in percent

REAGENT LEVEL

40155

Unsigned
Integer

1

R

0/100

the level of reagent in percent

Table 8 Sensor modbus registers

Tag Name

Register #

Data
Type

Length R/W

Discrete
Range

Min/Max
Range

Description

Advertising