Sensor standardization
Operating modes
• Entry of cell factor with simultaneous display
of conductivity and temperature
• Entry of conductivity of calibration solution
with simultaneous display of cell factor and
temperature
• Product calibration
• Zero adjustment
• Temperature probe adjustment
Permissible cell factor
00.100 ... 19.999
Permissible transfer ratio
01.00 ... 199.99
Permissible zero offset
± 0.5 mS/cm
Sensor monitoring
Sensocheck
• Monitoring of primary and wiring for
short circuit
• Monitoring of secondary and wiring for
open circuit
Sensoface
Provides information on the sensor condition
(evaluation of zero point, Sensocheck)
Sensor monitor
Sensor monitor for validation of sensor and
complete measured-value processing
(Display: resistance / temperature)
Temperature input
*
Pt100/Pt1000/NTC 100 kohms/NTC 30 kohms
2-wire connection, adjustable
Measuring range
Pt100 / Pt1000
-20 .. +200 °C (-4 ... 392 °F)
NTC100 kohm
-20 ... +130 °C (-4 ... 266 °F)
Resolution
0.1 °C / 1 °F
Measurement error
1,2,3)
0.5 K
(<1 K for Pt100; <1 K for NTC >100 °C)