Sensor standardization

Operating modes

• Entry of cell factor with simultaneous display

of conductivity and temperature

• Entry of conductivity of calibration solution

with simultaneous display of cell factor and

temperature

• Product calibration

• Zero adjustment

• Temperature probe adjustment

Permissible cell factor

00.100 ... 19.999

Permissible transfer ratio

01.00 ... 199.99

Permissible zero offset

± 0.5 mS/cm

Sensor monitoring

Sensocheck

• Monitoring of primary and wiring for

short circuit

• Monitoring of secondary and wiring for

open circuit

Sensoface

Provides information on the sensor condition

(evaluation of zero point, Sensocheck)

Sensor monitor

Sensor monitor for validation of sensor and

complete measured-value processing

(Display: resistance / temperature)

Temperature input

*

Pt100/Pt1000/NTC 100 kohms/NTC 30 kohms

2-wire connection, adjustable

Measuring range

Pt100 / Pt1000

-20 .. +200 °C (-4 ... 392 °F)

NTC100 kohm

-20 ... +130 °C (-4 ... 266 °F)

Resolution

0.1 °C / 1 °F

Measurement error

1,2,3)

0.5 K

(<1 K for Pt100; <1 K for NTC >100 °C)