Information on Calibration

Calibration can be performed by:

• Entry of cell factor

• Determining the cell factor with a known calibration solution taking

account of the temperature

• Product calibration

• Zero calibration in air or with calibration solution

• Temperature probe adjustment

Note:

If measurements are performed in fittings with A < 110 mm,

be sure to choose a calibration beaker with the same cross-

section and of the same material (metal/plastic).

Important Note:

All calibration procedures must be performed by trained person-

nel. Incorrectly set parameters may go unnoticed, but change the

measuring properties.

When another sensor is used, its sensor data (cell factor, transfer ratio,

measuring frequency, temperature probe) must be entered in the

configuration menu before calibration.

Each time a new sensor is connected, the transmitter must be

calibrated.

Exit

Press cal to exit. The measured

value is shown in the main

display alternately with “Hold”,

“enter” blinks, Sensoface is

active. Press enter key to end

the Hold mode. The measured

value is dispålayed. The output

current remains frozen for

another 20 sec (HOLD icon on,

“hourglass” blinks).