Alarm settings, Configuration, Exit: press conf , then enter – Hach-Lange SI794 I User Manual User Manual

Page 60

Advertising
background image

60

Error messages can also be signaled by a 22 mA output

current (see Pg 51,

The operating behavior of the alarm contact is shown on

Pg 98.

The alarm delay acts on the LED, the 22 mA signal, and

the alarm contact.

57, 96).

Configuration

Alarm settings

Select Sensocheck

Delay

LED in HOLD mode

Code Display

Menu group

Select menu item

Alarm settings

AL .

Exit:

Press conf, then enter.

15

16

Alarm

Alarm contact

The alarm contact is closed during normal operation (N/C).

It opens in the case of alarm or power outage.

As a result, a failure message is provided even in the case

of line breakage (fail-safe behavior). For contact ratings,

see Specifications.