Hach-Lange SI794 I User Manual User Manual

Page 83

Advertising
background image

83

Display

Action

Remark

Measuring mode:

From the blinking CAL mode

indicator you see that product

calibration has not been ter-

minated.

Product calibration step 2:

When the sample value has

been determined, open the

product calibration once more

(cal, code 1105).

Display (approx.

3 sec)

Enter sample value. The new

cell factor is calculated.

Press enter to confirm.

Until the sample

value is deter-

mined and can

be entered, the

transmitter is in

measuring mode.

The new cell factor and zero

point are displayed.

Press enter to confirm.

The measured value is shown

in the main display alternately

with “Hold”. “enter” blinks.

Press enter to exit.

After end of

calibration, the

outputs remain

in Hold mode for

approx. 20 sec.

New calibration:

Press cal.

Advertising