Code display action choices rl – Hach-Lange SI794 I User Manual User Manual
Page 63

63
LiMIT
(LiMIT
CtROL)
Code Display
Action
Choices
rL .
Use of relays:
• Limit function (LiMIT)
• Controller (CtROL)
Select using 4key.
Press enter to proceed.
Note: Selecting CtROL leads to
Controller menu Ct .
Lo
(Lo/Hi)
For limit 1 function principle,
see Pg 65. Select using 4 key.
Press enter to proceed.
Limit 1 contact response
N/O: normally open contact
N/C: normally closed contact
Select using 4key.
Press enter to proceed.
N/O
(N/O
N/C)
Limit 1 setpoint
Select position using 4key, edit
number using 5 key. Press enter to
proceed.
000 .0 mS
(xxxx)
Limit 1 hysteresis
Select position using 4key, edit
number using 5 key. Press enter to
proceed.
001 .0 mS
(xxxx)
Limit 1 delay
The contact is activated with delay
(deactivated without delay)
Select position using 4key,
edit number using 5 key.
Press enter to proceed.
0010 sec
(0 ...
9999 sec)
L1 .