Product calibration – Hach-Lange SI794 I User Manual User Manual

Page 82

Advertising
background image

82

The measurement process is only interrupted briefly. During

product calibration the sensor remains in the process.

Calibration is without TC correction!
Procedure: During sampling the currently measured value is

stored in the transmitter. The transmitter immediately returns

to measuring mode. The calibration mode indicator blinks and

reminds you that calibration has not been terminated. The sam-

ple is measured in the lab or directly on the site using a portable

meter. The measured sample value is then entered in the trans-

mitter. The new cell factor is calculated from these two values.

If the sample is invalid, you can take over the value stored dur-

ing sampling. In that case the old calibration values are saved.

Afterwards, you can start a new product calibration.

Display

Action

Remark

Product calibration step 1:

Press cal, enter code: 1105.

(Press 4 key to select position,

enter number using 5 key.

Press enter to confirm.)

Take sample and store value.

Press enter to proceed.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Now the sample

can be measured in

the lab.

Product calibration

Calibration by sampling

Display (approx.

3 sec)