Code display action choices o1, Concentration curves / ranges see page 112 et seq – Hach-Lange SI794 I User Manual User Manual

Page 43

Advertising
background image

43

Code Display

Action

Choices

o1 .

Only with 00.00 % Conc can you

select the process solution.

Select using 4 key.

-01- NaCl (0 ... 26 % by wt)
-02- HCl

(0 ...18 % by wt)

-03- NaOH (0 ... 14 % by wt)
-04- H

2

SO

4

(0 ... 30 % by wt)

-05- HNO

3

(0 ... 30 % by wt)

-06- H

2

SO

4

(92 ... 99 % by wt)

-07- HCl

(22 ... 29 % by wt)

-08- HNO

3

(35 ... 36 % by wt)

-09- H

2

SO

4

(32 ... 84 % by wt)

-10- NaOH (18 ... 50 % by wt)
Press enter to proceed.

-01-SOL

(-01-SOL

-02-SOL

-03-SOL

-04-SOL

-05-SOL

-06-SOL

-07-SOL

-08-SOL

-09-SOL

-10-SOL)

Concentration curves / ranges

see page 112 et seq.

The concentration curves of many sub-

stances show a maximum. This means that

if the substance concentration continues to

increase and the temperature remains con-

stant, the conductivity will drop. Therefore,

a one-to-one correlation of values is only

possible in defined ranges.

These partial ranges must be selected as

measurement ranges in the configuration.

Example:

Ranges for sulfuric acid