Code display action choices o1, Concentration curves / ranges see page 112 et seq – Hach-Lange SI794 I User Manual User Manual
Page 43

43
Code Display
Action
Choices
o1 .
Only with 00.00 % Conc can you
select the process solution.
Select using 4 key.
-01- NaCl (0 ... 26 % by wt)
-02- HCl
(0 ...18 % by wt)
-03- NaOH (0 ... 14 % by wt)
-04- H
2
SO
4
(0 ... 30 % by wt)
-05- HNO
3
(0 ... 30 % by wt)
-06- H
2
SO
4
(92 ... 99 % by wt)
-07- HCl
(22 ... 29 % by wt)
-08- HNO
3
(35 ... 36 % by wt)
-09- H
2
SO
4
(32 ... 84 % by wt)
-10- NaOH (18 ... 50 % by wt)
Press enter to proceed.
-01-SOL
(-01-SOL
-02-SOL
-03-SOL
-04-SOL
-05-SOL
-06-SOL
-07-SOL
-08-SOL
-09-SOL
-10-SOL)
Concentration curves / ranges
see page 112 et seq.
The concentration curves of many sub-
stances show a maximum. This means that
if the substance concentration continues to
increase and the temperature remains con-
stant, the conductivity will drop. Therefore,
a one-to-one correlation of values is only
possible in defined ranges.
These partial ranges must be selected as
measurement ranges in the configuration.
Example:
Ranges for sulfuric acid