Calibration by input of cell factor – Hach-Lange SI794 I User Manual User Manual
Page 78
Advertising
78
Calibration by input of cell factor
Input of cell factor with simultaneous display of conductivity and
temperature (without temperature compensation)
Display
Action
Remark
Enter cell factor:
Select position using 4 key,
edit number using5 key.
Conductivity and temperature
are alternately displayed dur-
ing the input (lower display).
Press enter to confirm entry.
Press cal key, enter code: 1100.
Select position using 4,
edit number using5,
press enter to proceed.
Ready for calibration
Remove and clean sensor
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
The entered cell factor and
zero point are displayed.
Press enter to confirm.
Display (3 sec)
Transmitter in Hold
mode, measured
value frozen.
Sensoface inactive.