Calibration by input of cell factor – Hach-Lange SI794 I User Manual User Manual

Page 78

Advertising
background image

78

Calibration by input of cell factor

Input of cell factor with simultaneous display of conductivity and

temperature (without temperature compensation)

Display

Action

Remark

Enter cell factor:

Select position using 4 key,

edit number using5 key.

Conductivity and temperature

are alternately displayed dur-

ing the input (lower display).

Press enter to confirm entry.

Press cal key, enter code: 1100.

Select position using 4,

edit number using5,

press enter to proceed.

Ready for calibration

Remove and clean sensor

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

The entered cell factor and

zero point are displayed.

Press enter to confirm.

Display (3 sec)

Transmitter in Hold

mode, measured

value frozen.

Sensoface inactive.