Zero calibration in air – Hach-Lange SI794 I User Manual User Manual

Page 84

Advertising
background image

84

Display

Action

Remark

When there has

not been an entry

for 6 sec, the lower

display alter-

nately shows the

conductivity and

temperature value.

Zero calibration in air

Press cal key, enter code: 1001.

Select position using 4,

edit number using5,

press enter to proceed.

Modify the zero point until

zero is displayed as conductiv-

ity value in the lower display.

Select using 4 key,

edit number using 5 key.

If required, change the sign of

the zero point.

Press enter to confirm the zero

point.

The transmitter is in

the Hold mode.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Ready for calibration

Remove and clean sensor.

(Sensor must be dry!)