Zero calibration with calibration solution – Hach-Lange SI794 I User Manual User Manual

Page 86

Advertising
background image

86

Display

Action

Remark

When there has

not been an entry

for 6 sec, the lower

display alternately

shows the conduc-

tivity and tempera-

ture value.

Zero calibration with calibration solution

Calibration solution with low conductivity

Press cal key, enter code: 1001.

Select position using 4,

edit number using5,

press enter to proceed.

Immerse sensor in calibration

solution.

Modify the value until the

lower display shows the

conductivity value of the

calibration solution.

Press enter to confirm

calibration.

The transmitter is in

the Hold mode.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Ready for calibration

Remove and clean sensor

Display (3 sec)

The cell factor and zero point

are displayed.

Press enter to confirm the

calibration data.