Zero calibration with calibration solution – Hach-Lange SI794 I User Manual User Manual
Page 86

86
Display
Action
Remark
When there has
not been an entry
for 6 sec, the lower
display alternately
shows the conduc-
tivity and tempera-
ture value.
Zero calibration with calibration solution
Calibration solution with low conductivity
Press cal key, enter code: 1001.
Select position using 4,
edit number using5,
press enter to proceed.
Immerse sensor in calibration
solution.
Modify the value until the
lower display shows the
conductivity value of the
calibration solution.
Press enter to confirm
calibration.
The transmitter is in
the Hold mode.
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Ready for calibration
Remove and clean sensor
Display (3 sec)
The cell factor and zero point
are displayed.
Press enter to confirm the
calibration data.