Diagnostics functions

Display of output currents

Press enter while in measuring mode.

The current at output 1 is shown in the main display,

the current at output 2 in the secondary display.

After 5 sec the transmitter returns to measuring

mode.

Display

Remark

Display of calibration data (Cal Info)

Press cal while in measuring mode and confirm code

0000. The current cell factor is shown in the main

display, the zero point underneath. After 20 sec the

transmitter returns to measuring mode (immediate

return at pressing enter).
Sensor monitor for validation

of sensor and complete measured-value processing.

Loop a defined sensing resistor (e.g. R = 100 Ω) through

the sensor as shown in the figure.

Press the conf key and enter code: 2222. The sensor

monitor displays the directly measured resistance and

the temperature.

If there is a significant difference between resistor

value and display, the sensor and its transmission

behavior should be checked.

Press enter to return to measurement.

Caution: The transmitter does not automatically go to

Hold mode.

Display of last error message

(Error Info)

Press conf while in measuring mode and confirm

code 0000. The last error message is displayed for

approx. 20 sec.

After that the message will be deleted (immediate

return to measurement at pressing enter).