Hach-Lange SI794 I User Manual User Manual

Page 39

Advertising
background image

39

Code Display

Action

Choices

o1 .

03 .000

• Select measuring frequency

Select using 4 key.

8 KHZ

(8 KHZ

12 KHZ)

100Pt

(100Pt

1000Pt

100NTC)

Press enter to proceed.

068 .80 Hz

• Select temperature probe

Select using 4 key.

Press enter to proceed.

With other selected

the sensor parameters are

entered separately:
• Enter cell factor:

(Select position using 4 key

and edit number using5 key.)

Press enter to proceed.

• Enter transfer ratio

Press enter to proceed.

Note

When other is called once more,

the last sensor parameters are

displayed and can be edited.

Note: Characters represented in gray are blinking and can be edited.

*) These parameters are only edited in parameter set 1.

They are valid for both parameter sets.