Calibration with calibration solution – Hach-Lange SI794 I User Manual User Manual

Page 80

Advertising
background image

80

Display

Action

Remark

Calibration with calibration solution

Press cal key, enter code: 0110.

Select position using 4,

edit number using5,

press enter to proceed.

Ready for calibration

Remove and clean sensor

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Display (3 sec)

Transmitter in Hold

mode, measured

value frozen.

Sensoface inactive.

Be sure to use known calibration solutions and the respective tempera-

ture-corrected conductivity values (see Calibration solutions Pg 106/107).

During the calibration procedure the temperature should be kept con-

stant.

Immerse sensor in calibration

solution.

Enter the temperature-

corrected conductivity value

of the calibration solution:

Select position using 4 key,

edit number using 5 key.

Cell factor and temperature

are alternately displayed in the

lower display.

Press enter to confirm entry.

When there has

not been an entry

for 6 sec, the lower

display alternately

shows the conduc-

tivity and tempera-

ture value.