Temp probe adjustment, Measurement – Hach-Lange SI794 I User Manual User Manual

Page 88

Advertising
background image

88

Display

Remark

Measurement

In the measuring mode the main display shows the

configured process variable (conductivity [mS/cm,

S/m], concentration [% by wt], or salinity [SAL]) and

the lower display shows the temperature.

The transmitter is switched to measuring mode

by pressing cal during calibration or by pressing

conf during configuration (waiting time for signal

stabilization approx. 20 sec).

Display

Action

Remark

Enter measured temperature

value.

Select position using4key,

edit number using 5key.

Press enter to proceed.

Press enter to exit adjustment.

HOLD will be deactivated after

20 sec.

Temp probe adjustment

Select calibration

(Press cal,

enter passcode: 1015)

Select position using4key,

edit number using 5key.

Press enter to proceed.

Measure the temperature of

the process medium using an

external thermometer

Wrong settings

change the measure-

ment properties! If

an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Default:

Value of secondary

display.

The transmitter is in

the Hold mode.