Temp probe adjustment, Measurement – Hach-Lange SI794 I User Manual User Manual
Page 88

88
Display
Remark
Measurement
In the measuring mode the main display shows the
configured process variable (conductivity [mS/cm,
S/m], concentration [% by wt], or salinity [SAL]) and
the lower display shows the temperature.
The transmitter is switched to measuring mode
by pressing cal during calibration or by pressing
conf during configuration (waiting time for signal
stabilization approx. 20 sec).
Display
Action
Remark
Enter measured temperature
value.
Select position using4key,
edit number using 5key.
Press enter to proceed.
Press enter to exit adjustment.
HOLD will be deactivated after
20 sec.
Temp probe adjustment
Select calibration
(Press cal,
enter passcode: 1015)
Select position using4key,
edit number using 5key.
Press enter to proceed.
Measure the temperature of
the process medium using an
external thermometer
Wrong settings
change the measure-
ment properties! If
an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Default:
Value of secondary
display.
The transmitter is in
the Hold mode.