Hach-Lange EVITA OXY Basic User Manual User Manual
Page 110

English 110
Reset
The system can be set back to the default values.
Figure 32 Reset
Send command
Via the "Send command" menu you can initiate the calibration process,
start the measurement procedure after the calibration, set the operating
hours counter and set the system to default settings.
Figure 33 Send command
Advanced settings
The "Advanced settings" menu enables the selection of calibration type
and the way in which the measurement procedure should be started
after the calibration. Furthermore factors for the compensation of the
salinity and of the relative humidity can also be entered.
"Automatic calibration" means calibration in atmospheric air and "manual
calibration" means calibration in a standard solution, whose oxygen
concentration can be entered via the keypad (refer to
The "transmitter address" must be for the output 0 in order to function in
a point-to-point installation with use of USC 5000. With operation of a
multidrop system with use of USC 7000 the transmitter must have a
clear address between 1 and 15.