Hach-Lange EVITA OXY Basic User Manual User Manual

Page 110

Advertising
background image

English 110

Reset

The system can be set back to the default values.

Figure 32 Reset

Send command

Via the "Send command" menu you can initiate the calibration process,
start the measurement procedure after the calibration, set the operating
hours counter and set the system to default settings.

Figure 33 Send command

Advanced settings

The "Advanced settings" menu enables the selection of calibration type
and the way in which the measurement procedure should be started
after the calibration. Furthermore factors for the compensation of the
salinity and of the relative humidity can also be entered.
"Automatic calibration" means calibration in atmospheric air and "manual
calibration" means calibration in a standard solution, whose oxygen
concentration can be entered via the keypad (refer to

Figure 27,

page 107

).

The "transmitter address" must be for the output 0 in order to function in
a point-to-point installation with use of USC 5000. With operation of a
multidrop system with use of USC 7000 the transmitter must have a
clear address between 1 and 15.