English 118

Flashing bell symbol
in the display

Not defined

WU10

I OUT 1: Values at 4 mA \>= values at 20 mA

Check setup of I OUT 1

WU11

I OUT 2: Values at 4 mA \>= values at 20 mA

Check setup of I OUT 2

WU12

I OUT 3: Values at 4 mA \>= values at 20 mA

Check setup of I OUT 3

WU13

I OUT 4: Values at 4 mA \>= values at 20 mA

Check setup of I OUT 4

WU14

Measurement value above current output setting 1 at 20 mA

Change values at 20 mA for I OUT 1

WU15

Measurement value above current output setting 2 at 20 mA

Change values at 20 mA for I OUT 2

WU16

Measurement value above current output setting 3 at 20 mA

Change values at 20 mA for I OUT 3

WU17

Measurement value above current output setting 4 at 20 mA

Change values at 20 mA for I OUT 4

WU18

Measurement value above current output setting 1 at 4 mA

Change values at 4 mA for I OUT 1

WU19

Measurement value above current output setting 2 at 4 mA

Change values at 4 mA for I OUT 2

WU20

Measurement value above current output setting 3 at 4 mA

Change values at 4 mA for I OUT 3

WU21

Measurement value above current output setting 4 at 4 mA

Change values at 4 mA for I OUT 4

WU39

HART data exchange error

Check cabling

WU40

TMS data exchange error

Replace TMS

WU41

A parameter is outside of the range.
Could not be replaced by the standard value.

Is remedied by new switch-on

WU43

Too many errors simultaneously.
Some errors are not correctly registered.

Switch USC off and on again

WU49

Internal error

Switch USC off and on again

One or several
outputs = 3.8 mA
or 24 mA

EU60

The CAN bus application is not responding.
This error is shown when the maximum number of
communication attempts has been exceeded.

1. Switch USC off and on again
2. Replace USC

Symptom

Current output

Event
code

Cause

Resolution