Hach-Lange EVITA OXY Basic User Manual User Manual

Page 119

Advertising
background image

English 119

Flashing bell symbol
in the display

Not defined

EU61

USC cannot read the data for current output
I OUT 1
1

No response from the transmitter

2

Transmitter has received a new address

3

Another master is calling up problems with the
HART

®

communication.

4

Supply voltage for transmitter
too low

5

Too much noise interference on the current loop

6

Transmitter defective

7

USC defective

1

Check cable and shielding

2

Switch USC off and on again

3

Disconnect other master, or set up
USC for other master, refer to figure
51, page 52

4

Check supply voltage of the transmit-
ter; it must be at least 20 V.

5

Check cable and shielding

6

Replace transmitter

7

Replace USC

EU62

USC cannot read the data for current output I OUT 2
1–7: As with EU61

As with

EU61, page 117

EU63

USC cannot read the data for current output I OUT 3
1–7: As with EU61

EU64

USC cannot read the data for current output I OUT 4
1–7: As with EU61

EU65

USC cannot read the data for relay 1
As with EU61

EU66

USC cannot read the data for relay 2
As with EU61

EU67

USC cannot read the data for relay 3
As with EU61

EU68

USC cannot read the data for relay 4
As with EU61

EU69

USC cannot read the data for relay 5
As with EU61

Symptom

Current output

Event
code

Cause

Resolution

Advertising