Guide to the troubleshooting – transmitter

OXY 4100/3150/4150

Symptom

Output signal

Event
code

Cause

Resolution

Incorrect measurement
value

Variable

Incorrect calibration value

1

Check full-scale value in atmospheric air. Hold the
transmitter in the open air with the sensor down-
ward. Compare the value with DS/EN 25814 (refer
to appendix IIa)

2

Check the zero point (refer to appendix IIb)

3

If significant deviation, calibrate again

Display switches between
"Oxygen" and "End of
calibration"

End of the calibration phase

Switch to the "Send command" menu, and begin
measurement, refer to

Figure 33, page 110

Flashing bell in the
display field

Depending on
settings

WO10

Measurement value above the set
transmitter output setting

Change upper range under "Basic settings",
refer to

Figure 21, page 102

WO11

Measurement value under the set
transmitter output setting

Change lower range under "Basic settings",
refer to

Figure 22, page 102

WO15

Service life expired. OXY 1100 should be
replaced as quickly as possible

Replace OXY 1100 sensor

WO20

Temperature of OXY 1000 below 0 °C.
Accuracy potentially reduced.

Inaccuracy

WO21

Temperature of OXY 1000 above 40 °C.
Accuracy potentially reduced.

Inaccuracy

EO50

Calibration not possible.
OXY 1100 worn out or defective

Replace OXY 1100 sensor

EO51

Output signal of OXY 1100 too weak.
OXY 1100 worn out or defective

Replace OXY 1100 sensor

EO55

The OXY 1100 membrane has become
damaged

Replace OXY 1100 sensor

EO60

EPROM error

Replace transmitter

EO61

Transfer error between μ-controller and A/D
converter
or error in the A/D converter

Replace transmitter

EO62

Error in the input cycle

Replace transmitter

EO63

Error in the temperature cycle

Replace transmitter