Mechanical installation – Hach-Lange EVITA OXY Basic User Manual User Manual

Page 88

Advertising
background image

English 88

Figure 2 EVITA system

The EVITA

®

OXY system offers many advantages:

Transmitter with two wire technology

effortless and fast installation

allows up to 1000 m between the transmitter and a USC signal
converter, a display or a PLC

Problem-free maintenance

problem-free, automatic calibration process in atmospheric air
through the TILTCAL

®

device: simply hold the transmitter for

5 minutes in tilted position

the float ball is self-cleaning

after 2–3 years the Clark sensor is easily replaced

High reliability

EVITA

®

OXY features a self-diagnosis device

the membrane is permanently checked for leak points

the fault reporting is performed via the analog output of the
transmitter

Flexible measurement value transfer

The standard HART

®

communication offers additional possibilities:

remote adjustment of the measurement range and of the

measurement units

display of the temperature and of the volume of dissolved

oxygen, the still remaining life of the sensor OXY 1100 and

special event codes

Tool-free insertion of the optionally available additional module into
the USC 7000.

Mechanical installation

For wall installation: Screws/hardware corresponding to the wall
properties and with sufficient load-bearing capacity must be used.
Maintenance- and repair work should be carried out only by the
authorized customer service. Only original replacement- and accessory
parts recommended by the manufacturer may be used. Changes to the
product have the loss of any liability claim as the result.

Transmitter

Installation on the handrail
The holders for the transmitters OXY 4100 and 4150 are attached with
help of the supplied hose clamps directly to the handrail. The hose
clamps must be tightened after a few days to make sure of the firm seat
of the holder.
If strong lateral forces exist, the PVC pipe must also be attached with
support wires. In this way the lateral forces working on the pipe are
reduced and a breakage of the holder prevented.

N O T I C E

The installation must be performed by a specialist (qualified personnel) under
compliance with all locally valid safety regulations. The installation site is to be
chosen so that a safe installation and the safe operation is guaranteed.