Guide to the troubleshooting – usc – Hach-Lange EVITA OXY Basic User Manual User Manual

Page 117

Advertising
background image

English 117

Guide to the troubleshooting – USC

Symptom

Current output

Event
code

Cause

Resolution

None

Not defined

IU00

Switch-on

Registration of the switch-on time

IU01

Additional module installed

Registration of the installed
additional module

IU02

Additional module defective or removed

Replace or install additional module

IU03

Parameter successfully corrected

Information

No display

1

Supply voltage missing/too low

2

USC 5000/7000 defective

1

Check supply voltage

2

Replace USC 5000/7000

Dots in the uppermost
line of the display

(I OUT 1)

EU61

Refer to

EU61, page 117

As with

EU61, page 117

(I OUT 2)

EU62

Refer to

EU62, page 117

(I OUT 3)

EU63

Refer to

EU63, page 117

(I OUT 4)

EU64

Refer to

EU64, page 117