background image

English 85

Ambient temperature

Storage:
Operation:

-40° C to + 70° C
-40° C to + 50° C

Storage:
Operation:

-40° C to + 70° C
-40° C to + 50° C

Environment

Indoor use, height up to 2000 m
Pollution degree II,
Relative humidity 80 % up to 31 °C
Linear decreasing to 50 % at 40 °C

Indoor use, height up to 2000 m
Pollution degree II,
Relative humidity 80 % up to 31 °C
Linear decreasing to 50 % at 40 °C

Power supply

100–240 Vac +10 %/–15 %,
50/60 Hz. 5–20 VA
Fuse: T500 mA/250 V
(not to be changed by the user)

100–240 Vac +10 %/–15 %,
50/60 Hz. 5–20 VA
Fuse: T500 mA/250 V
(not to be changed by the user)

11–30 VAC/DC 5-12 W

Automatic
calibration

Compensation of the temperature, of the air pressure,
of the humidity and of the salinity

Compensation of the temperature, of the air pressure,
of the humidity and of the salinity

Cable gland

Jacob GmbH, catalogue number 50.013 PA (PG 13.5)
6–12 mm cable diameter

Jacob GmbH, catalogue number 50.013 PA (PG 13.5)
6–12 mm cable diameter

Blind screw connection

Jacob GmbH, catalogue number 1013 PA

Jacob GmbH, catalogue number 1013 PA

Approvals

Approval according to CE
Emission/immunity EMC: EN 61326

Approval according to CE
Emission/immunity EMC: EN 61326

* The insulation between the mains connection and all in- and output terminals of the USC signal converter
is provided by double or reinforced insulation of 2300 V AC strength (category II).

Important:
Before the return of goods to Hach Lange, the products must be thoroughly cleaned.