Appendix i appendix ii a appendix ii b, Warranty, liability and complaints – Hach-Lange EVITA OXY Basic User Manual User Manual

Page 121

Advertising
background image

English 121

Appendix I

Appendix II a

Appendix II b

Check of the zero point

Mix 1 g sodium sulfite (Na

2

SO

3

), 1 mg cobalt (II) salt, (cobalt (II) chloride

hexahydrate CoCI

2

, 6H

2

O) with 1 liter water to remove the oxygen from

the water. Then stir the water and measure the oxygen concentration to
determine the zero point.

Warranty, liability and complaints

Hach Lange GmbH warrants that the supplied product is free of
material- and processing defects and undertakes to repair or to replace
any defective parts at zero cost (no composition- or durability
guarantee).
Deviating from our general terms of business, a limitation period for
defect-related claims of 12 months applies for the devices of the EVITA
series.

Default values and significance of the deviation from these values

Parameter

Default
setting

Deviation

Significance (as % of
the current value)

Atmospheric
air pressure

1013.25 mb
ar

10 mbar

approximately 1

Relative
humidity

100 %

10 %

approximately 0.3

Salinity

0 g/L

1 g/L

approximately 1

Temperature

Solubility of oxygen in water in equilibrium with air
(»100 %) at 101.325 kPa

°C

mg/L or ppm

0

14.62

1

14.22

2

13.83

3

13.46

4

13.11

5

12.77

6

12.45

7

12.14

8

11.84

9

11.56

10

11.29

11

11.03

12

10.78

13

10.54

14

10.31

15

10.08

16

9.87

17

9.66

18

9.47

19

9.28

20

9.09

21

8.91

22

8.74

23

8.58

24

8.42

25

8.26

26

8.11

27

7.97

28

7.83

29

7.69

30

7.56

Advertising