English 112 – Hach-Lange EVITA OXY Basic User Manual User Manual

Page 112

Advertising
background image

English 112

Monitoring of the current- and of the relay outputs

Figure 35 Monitoring of the current- and of the relay outputs

Service mode

In the service mode there is the option to monitor the relay
and the current outputs directly. This can be very helpful if
for example a control loop should be
set or an alarm signal should be checked.

The quality of the HART® communication is shown in the
sub menu: "HART comm. Error".

The quality is shown as follows:

VVVV XXXX YYY % ZZ

where:

VVVV is the number of the reports without error,

XXXX is the number of the reports with error,

YYY % is messages with errors in percent,

ZZ is the number of the reports that have been deleted
because for example the maximum number of repetitions
has been reached.

Service

mode

USC

I Out 2

USC

I Out 3

USC

I Out 4

USC

Relay 1

USC

Relay 2

USC

Relay 3

USC

Relay 4

USC

Relay 5

USC

Digital input

High

USC

(High/Low)

Mode

Normal

I OUT 1

Normal

Forced

Forced value

00.0 mA

I OUT 1

00.0-24.0 mA

Mode

Normal

RELAY 1

Normal

Open

Closed

Event log

USC

Log xxxxdxxh

xxxxxxxxxxxxxxxx

USC

I Out 1

USC

081R9406.02.02

HART comm. error

0856 0007 001% 00

USC